• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723RNAB
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had a blank display and stopped working a month ago.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed displacement test, self test, off no power test and all operating currents tested within the specification.No blank display were noted.The test reservoir p-cap was able to lock in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key10337025
Report Number2032227-2020-154153
Device Sequence Number3032303
Product Code OYC
UDI-Device Identifier00613994748713
UDI-Public(01)00613994748713
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723RNAB
Device Catalogue NumberMMT-723RNAB
Device Lot NumberA000167986
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2020
Initial Date Received by Manufacturer 07/23/2020
Supplement Date Received by Manufacturer08/14/2020
Initial Report FDA Received Date07/29/2020
Supplement Report FDA Received Date08/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
Patient Weight79
-
-