Information received from initial reporter via medtronic field representative regarding an event happened during intra-op for pre-operative diagnosis of degenerative disc disease.It was reported that crosslink was broken and the set screws of the crosslink were sheared off and broken.It was mentioned that upon torquing the set screw, they noticed that the set screw had broke and the bottom portion of the set screw was sheared off.The set screw was replaced and a second attempt of torquing took place, again the set screw sheared off.Later, they removed entire crosslink and replaced a new crosslink of same size and then torquing was successful.All the broken parts of crosslink and setscrew were removed.It was confirmed that there were no fragments of implant/instrument left inside the patient.There were no symptoms reported due to this event.There were no complications to patient/physician reported/anticipated.
|
|
H3: product analysis of product# 8115545 lot # 0672033w visually confirmed the set screw is fractured approximately ~1 thread up from the distal tip of the set screw.Microscopic examination of the fracture surface appears to emanate from the root of the thread, following the thread helix, with downward deformation of the sheared portion of the thread.The outer hex does not appear to be damaged however, the inner hex does show signs of deformation.The above observations are consistent with torsional shearing of the thread during tightening.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|