Product analysis: it was also determined at analysis that the programmer suddenly displayed a warning message ¿programmer overheating please turn the programmer off¿, this was due to a worn power supply.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
It was reported the patient is pacemaker dependent, during the procedure whilst using the analysers surgical cable, transmission was lost between the analyser and the implanted electrode.There was a sudden loss of stimulation, no peaks were visible on the monitor, connection and the lead position was checked, cardiac massage during asystole was required.After change to another programmer and another measuring cable regular stimulation took place again.The procedure was completed using a different device.It was further reported the programmer was used the following day with no issues reported.Hospitalisation and intervention was required as a result of this event.The programmer will be returned for service to be tested with the cable.Injury reported.
|