Date of Event 06/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the second unit of the double red cell collection could be due to the flow not going through the filter medium as expected.The centrifuge pressure was lower during the collection of the second unit, and the overall product volume of the second unit is higher than expected.This could be due to a faulty filter, filter misload, or if the filter had been removed from the bracket at any time during collection.Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the red blood cell (rbc) product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(6).The collection set is not available for return because it was discarded by the customer.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Root cause: the signals in the run data file indicate that the higher-than-expected wbc content in the second unit of the double red cell collection could be due to the flow not going through the filter medium as expected.The centrifuge pressure was lower during the collection of the second unit, and the overall product volume of the second unit is higher than expected.This could be due to a faulty filter, filter misload, or if the filter had been removed from the bracket at any time during collection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA ACCEL RBC PLS AUTORAS
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key10338312
Report Number1722028-2020-00354
Device Sequence Number19301540
Product Code GKT
UDI-Device Identifier05020583805200
UDI-Public05020583805200
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
BK190332
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2022
Device Model Number80520
Device Catalogue Number80520
Device Lot Number2005212230
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2020
Supplement Date Received by Manufacturer01/26/2021
Initial Report FDA Received Date07/29/2020
Supplement Report FDA Received Date02/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-