• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550275-33
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intimal Dissection (1333)
Date of Event 07/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.It should be noted that the reported patient effect of intimal dissection is listed in the xience sierra everolimus eluting coronary stent system instructions for use as known patient effect of coronary stenting procedures.Based on the information reviewed, there is no indication of a product quality issue.A conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.The additional xience sierra device referenced is filed under a separate medwatch report number.
 
Event or Problem Description
Patient id: (b)(6).It was reported that on (b)(6) 2020, the patient presented with a recent non-st elevated myocardial infarction (nstemi), elevated troponin, and a percutaneous coronary intervention was performed.Pre-dilatation was performed on the mid left anterior descending (lad) coronary artery lesion and a 2.75x33mm xience sierra stent (1550275-33, 0011041) was implanted.An edge dissection was observed, treated with a 3.0x33mm xience sierra stent (1550300-33, 0021841).Post-dilatation was performed on the 3.0x33mm xience sierra stent when a plaque shift with 100% occlusion occurred in the proximal diagonal coronary artery.Additional angioplasty was performed as treatment.Reportedly, there was no device malfunction and the event resolved.Post-procedure results were acceptable, 0% residual diameter stenosis and timi flow iii.No additional information was provided regarding this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10338703
Report Number2024168-2020-06278
Device Sequence Number14223970
Product Code NIQ
UDI-Device Identifier08717648227219
UDI-Public08717648227219
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2021
Device Model Number1550275-33
Device Catalogue Number1550275-33
Device Lot Number0011041
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2020
Initial Report FDA Received Date07/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient Weight102
-
-