• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX EMBOLIZATION DEVICE; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX EMBOLIZATION DEVICE; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED-375-16
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the pipeline was implanted without issue.However, upon retrieving the delivery system, the tip coil broke off and had to be snared.Post-procedure angiographic results showed the implant was fine, and there was no impact to the patient.It was indicated that all devices were prepared as per the instructions for use (ifu).The patient was undergoing surgery for an off-label treatment of an sah and a ruptured aneurysm.
 
Additional Manufacturer Narrative
B5: event description - additional information d4: suspected medical device - model, lot number, expiration date, unique identifier (udi) #: correction h4: device mfg date - correction medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received some additional information.The vessel was normal tortuous.Aneurysm located in supraclinoid internal carotid artery.No additional resistance prior to break.
 
Additional Manufacturer Narrative
No bends or kinks were found with the pipeline flex pushwire.The pipeline flex embolization device was found be broken at distal end of hypotube.The ptfe shrink tubing was found to be damaged on proximal broken segment.The proximal bumper, pad and re-sheathing marker were found intact.The dps restraints/sleeves were found to be intact.The tip coil was found to be damaged.The broken distal hypotube was sent out for sem (scanning electron micrographic) / eds (energy dispersive spectroscopy) elemental analysis.The elemental analysis of the broken hypotube end shows tin (sn) was detected on the surface.No other anomalies were observed.There was an indication that the event could be related to a potential manufacturing issue; therefore, a device history record review will be performed.Based on the device analysis and reported information, the customer¿s report of ¿pushwire break/separation¿ was confirmed.Possible contributors towards failure are patient vessel tortuosity, or user re-sheaths more than two times.Customer reported patient vessel tortuosity as normal.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX EMBOLIZATION DEVICE
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key10338748
Report Number2029214-2020-00743
Device Sequence Number16885154
Product Code OUT
UDI-Device Identifier00847536015661
UDI-Public00847536015661
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P100018.S011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model NumberPED-375-16
Device Catalogue NumberPED-375-16
Device Lot NumberA823545
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2020
Initial Date Received by Manufacturer 07/27/2020
Supplement Date Received by Manufacturer07/30/2020
11/05/2020
Initial Report FDA Received Date07/29/2020
Supplement Report FDA Received Date08/17/2020
11/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-