| Model Number |
9-PDA-003 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Codes |
Embolism (1829); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, an amplatzer duct occluder (9-pda-003) was selected for implant in the patient using a 5f amplatzer torqvue delivery system.During the procedure, the delivery system was inserted into the patient, but it was impossible to advance the amplatzer duct occluder onto the delivery system.There was resistance felt at the connection with the loading tube and the initial part of the delivery system.The physician decided to exchange the delivery system for a 6f amplatzer torqvue delivery system.There was no resistance felt and the amplatzer duct occluder was successfully implanted in the patient with no significant delay in the procedure.A few seconds after the device was placed in the patient, the device embolized to the left iliac artery of the patient.The device was snared through a 10f introducer with no patient complications.Immediately after the device was removed, the physician implanted a larger amplatzer duct occluder (9-pda-005) with the 6f amplatzer torqvue delivery system with no complications.The patient experienced no serious injuries and is currently discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
The report was submitted under the incorrect product code.The correct product code is mae.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of difficulty advancing the occluder through the sheath and device embolization was reported.The investigation confirmed the device met dimensional and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
|
| |
|