• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SROM NECK TRL 36STD +12LAT; HIP INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SROM NECK TRL 36STD +12LAT; HIP INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Model Number 2576-36-012
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Code Available (3191)
Date of Event 07/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the 36+12 s-rom neck segment not locking down onto the trial stem.No surgical delay.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.¿.
 
Additional Manufacturer Narrative
Product complaint #
=
>(b)(4) investigation summary
=
> examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.H10 additional narrative: added: d4 (lot) corrected: g1, h3, h8.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SROM NECK TRL 36STD +12LAT
Common Device Name
HIP INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10339863
Report Number1818910-2020-17165
Device Sequence Number14211158
Product Code LXH
UDI-Device Identifier10603295147084
UDI-Public10603295147084
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2576-36-012
Device Catalogue Number257636012
Device Lot NumberAF1003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2020
Initial Date Received by Manufacturer 07/21/2020
Supplement Date Received by Manufacturer08/15/2020
08/20/2020
Initial Report FDA Received Date07/29/2020
Supplement Report FDA Received Date08/19/2020
08/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-