| Model Number |
D002 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Battery Problem (2885)
|
| Health Effect - Clinical Codes |
No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/05/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product has been received for analysis.This report will be updated upon completion of analysis.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) declared end of life (eol) and was in safety mode following abdominal surgery underwent by this patient.During a previous follow up in (b)(6) 2019, the device displayed normal battery depletion.Subsequently, this icd was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) declared end of life (eol) and was in safety mode following abdominal surgery for this patient.During a previous follow up in november 2019, the device displayed normal battery depletion.The device was explanted and successfully replaced.No additional adverse patient effects were reported.This device has been returned and analysis has been completed.This report has been updated with the product investigation results.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned implantable cardioverter defibrillator (icd) was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, defibrillation, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
|
| |
|