• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ORIGEN MINI ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ORIGEN MINI ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D002
Medical Device Problem Codes Premature Discharge of Battery (1057); Battery Problem (2885)
Health Effect - Clinical Codes No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product has been received for analysis.This report will be updated upon completion of analysis.(b)(4).
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) declared end of life (eol) and was in safety mode following abdominal surgery underwent by this patient.During a previous follow up in (b)(6) 2019, the device displayed normal battery depletion.Subsequently, this icd was explanted and replaced.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) declared end of life (eol) and was in safety mode following abdominal surgery for this patient.During a previous follow up in november 2019, the device displayed normal battery depletion.The device was explanted and successfully replaced.No additional adverse patient effects were reported.This device has been returned and analysis has been completed.This report has been updated with the product investigation results.
 
Additional Manufacturer Narrative
The returned implantable cardioverter defibrillator (icd) was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, defibrillation, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORIGEN MINI ICD DR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key10340876
Report Number2124215-2020-15954
Device Sequence Number3178560
Product Code LWS
UDI-Device Identifier00802526533747
UDI-Public00802526533747
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/26/2016
Device Model NumberD002
Device Catalogue NumberD002
Device Lot Number102009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2020
Initial Date Received by Manufacturer 06/05/2020
Supplement Date Received by Manufacturer02/05/2021
Initial Report FDA Received Date07/29/2020
Supplement Report FDA Received Date02/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
-
-