SMITH & NEPHEW, INC. HEALICOIL RG SA 5.5MM W/2 UB-BL CBRD BL; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72203706
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/17/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the surgery it was reported that the packaging was opened, as the device was no longer sterile it was not used.No patient injuries or delay reported.Smith and nephew back-up device was available to complete the surgery.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
One 72203706 healicoil regenesorb 5.5mm suture anchor intended for use in treatment, was returned.The foil pouch was open on the straight seal edge.There was no evidence of a previous seal whatsoever as clearly seen with the side edges.If it had been sealed, there would be visible confirmation.Having a missing seal is not the intended condition for our distribution.Smith and nephew instructs return of product as sold when an observance is noted.The product was received in as sold condition with exception of the pouch seal.Complaint history review indicated no similar allegations for the lot number reported.Batch review did not indicate a condition, product or procedure failure that supported the allegation, although this product was part of a quarantine for a systemic absent seal issue.The failure was found post partial lot release to distribution.A general evaluation for absent seals was conducted across product codes.This lot was manufactured during that timeframe and was captured on the affected product list.No additional actions required at this time.Quarantine actions are now complete.Occurrences are currently being monitored by surveillance.
 
Additional Manufacturer Narrative
H3, h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.Visual assessment showed the foil pouch was open on the straight seal edge.There was no evidence of a previous seal whatsoever as clearly seen with the side edges.If it had been sealed, there would be visible confirmation.A product specification review indicated the findings were a defect.A complaint history review indicated no similar allegations for the lot number reported.Batch review did not indicate a condition, product or procedure failure that supported the allegation, although this product was part of a quarantine for a systemic absent seal issue identified post partial lot release.The complaint was confirmed, and the root cause was determined to be a manufacturing error.A corrective action has been initiated to mitigate future recurrence of similar events.Information corrected in d9, h6, h7, h9 and h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALICOIL RG SA 5.5MM W/2 UB-BL CBRD BL
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10342008
Report Number1219602-2020-01152
Device Sequence Number16358630
Product Code MAI
UDI-Device Identifier00885554031416
UDI-Public00885554031416
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/15/2023
Device Model Number72203706
Device Catalogue Number72203706
Device Lot Number2048395
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Initial Date Received by Manufacturer 07/17/2020
Supplement Date Received by Manufacturer09/04/2020
10/18/2021
Initial Report FDA Received Date07/29/2020
Supplement Report FDA Received Date09/11/2020
10/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0136-2021
Patient Sequence Number1
-
-