MEDTRONIC SOFAMOR DANEK USA, INC KYPHX HV-R BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Model Number CX01A
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes Extravasation (1842); Pain (1994)
Date of Event 07/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare professional via manufacturer representative regarding as patient having kyphoplasty due osteoporotic fracture of l4.It was reported that cement extravasation occurred during filling of the l4 vertebral body on the left side and extended into the psoas muscles at the mid body position.Cement was at the 7 minute post mix time point.Post procedure patient complained of left sided groin and thigh pain and was admitted for pain control.It was reported that cement was mixed for 25 seconds.The product was implanted on (b)(6) 2020 and remains implanted.It was reported that patient pain has been resolved and there are no further symptoms.There were no further complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHX HV-R BONE CEMENT
Common Device Name
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key10343244
Report Number1030489-2020-01007
Device Sequence Number16344609
Product Code NDN
UDI-Device Identifier00643169097803
UDI-Public00643169097803
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K102397
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberCX01A
Device Catalogue NumberCX01A
Device Lot NumberEL70180
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2020
Initial Report FDA Received Date07/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age74 YR
Patient Weight92
-
-