| Model Number |
21CAVGJ-514 00 |
| Medical Device Problem Code |
Biocompatibility (2886)
|
| Health Effect - Clinical Code |
Thrombus (2101)
|
| Date of Event |
07/23/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
In 2016, a 21mm masters series cavgj mechanical valve was implanted.On (b)(6) 2020, the valve was explanted due to thrombus that was unable to be removed via thromboembolisation.A competitor's 23mm sorin conduit valve was successfully implanted.The patient is reported to be in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Explant was reported due to thrombus.The investigation found that both mechanical leaflets were dislodged from the orifice.One was fractured.There was a tiny piece of white material within on recessed pivot area, which was too small to sample.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
|
| |
|