• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMMUCOR, INC. ECHO; AUTOMATED BLOOD BANK SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMMUCOR, INC. ECHO; AUTOMATED BLOOD BANK SYSTEM Back to Search Results
Model Number 0087000
Medical Device Problem Codes False Negative Result (1225); False Positive Result (1227)
Health Effect - Clinical Code Test Result (2695)
Type of Reportable Event Malfunction
Event or Problem Description
This report summarizes malfunction events.A review of the events indicated that three (3) patient sample tests using the echo v2 automated blood bank system produced unexpected negative results for specific antibodies and for rh typing expectations.The results represent false negatives due to prior patient histories or, testing completed by other methods that demonstrated the presence of the specific antibodies.The malfunction did not show positive results as expected for the anti-e antibody and the anti-k antibody on one (1) occasion when using the capture r ready id reagent.The malfunction did not show produce rh typing results as expected on two (2) occasions.Subsequent testing of reagent retention lots, reviews of design history records and reagent antigen validation testing of the initial bulk product used for commercial vialing showed acceptable results.Through post event tests and investigations, no specific causes were determined for the malfunctions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHO
Common Device Name
AUTOMATED BLOOD BANK SYSTEM
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross, ga
MDR Report Key10345948
Report Number1034569-2020-00040
Device Sequence Number2941936
Product Code KSZ
UDI-Device Identifier10888234001584
UDI-Public10888234001584
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
BK070016
Number of Events Summarized3
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number0087000
Device Catalogue Number0087000
Device Lot Number492400,ID398
Was Device Available for Evaluation? Yes
Type of Report(Section G)Periodic
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-