• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNITED STATES ENDOSCOPY GROUP, INC. RAPTOR GRASPING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNITED STATES ENDOSCOPY GROUP, INC. RAPTOR GRASPING DEVICE Back to Search Results
Model Number 00711177
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device subject of this event has not been returned to steris endoscopy for evaluation; however, the distributor has provided photographs of the device and a video of the device being actuated.Contrary to the initial report, examination of photographs and video of device actuation has found no evidence of damage or malfunction.The device history record was reviewed and confirmed the device lot was manufactured to specification.There have been no other complaints associated with this lot.The instructions for use include the following statements: "actuate the device by moving the slider on the handle back and forth to confirm that the grasping jaws open and close smoothly.If the unit does not function properly, or there is evidence of damage (e.G.Bends, kinks, misshapen jaws, misaligned jaws, exposed wires) do not use this product and contact your local product specialist." steris endoscopy has requested that the distributor offer in-service training to the user facility.No further issues have been reported.This report will be updated if additional information becomes available.
 
Event or Problem Description
The distributor reported that a raptor grasping device was damaged prior to use and could not be closed.The device was not used in a procedure nor inserted into the endoscope.The patient had no contact with the device and there was no harm to the patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPTOR GRASPING DEVICE
Common Device Name
GRASPING DEVICE
Manufacturer (Section D)
UNITED STATES ENDOSCOPY GROUP, INC.
5976 heisley rd
mentor, oh
MDR Report Key10346007
Report Number1528319-2020-00024
Device Sequence Number3447466
Product Code OCZ
UDI-Device Identifier00816765011317
UDI-Public(01)00816765011317
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K120084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2022
Device Model Number00711177
Device Catalogue Number00711177
Device Lot Number1912102
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2020
Initial Report FDA Received Date07/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-