• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER; NRY Back to Search Results
Model Number 5MAXACE068
Medical Device Problem Code Stretched (1601)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra system ace 68 reperfusion catheter (ace68) and a penumbra system aspiration pump max 220 (pump max).During the procedure, while the pump max was powered on, the physician felt some shaking when the distal end of the ace68 moved on the monitor and subsequently, the physician decided to remove the ace68.Upon removal of the ace68, it was noticed that the distal end of the ace68 was stretched.Therefore, the ace68 was not used for the remainder of the procedure.The procedure was completed using a new ace68 and the same pump max.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the returned ace68 was kinked at approximately 128.0 cm, 129.0 cm, and 132.0 cm from the hub.The total length was measured approximately 133.0 cm from the hub.Conclusions: evaluation of the returned ace68 revealed kinks on its distal shaft.If the device is advanced against resistance, damage such as distal kinks may occur.During functional testing, water was aspirated through the ace68 and the reported shaking on the distal end was not observed.The total length of the ace68 was measured within specification; therefore, the reported distal end stretching could not be confirmed.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10346043
Report Number3005168196-2020-01052
Device Sequence Number824679
Product Code NRY
UDI-Device Identifier00814548016535
UDI-Public00814548016535
Combination Product (Y/N)Y
Initial Reporter CountryKS
PMA/510(K) Number
K161640
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/06/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/02/2022
Device Model Number5MAXACE068
Device Catalogue Number5MAXACE068
Device Lot NumberF90059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/06/2020
Supplement Date Received by Manufacturer08/05/2020
Initial Report FDA Received Date07/30/2020
Supplement Report FDA Received Date09/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
Date Report Sent to FDA01/01/2005
-
-