|
(b)(4).Batch #: m53v0d.Device analysis: the analysis results found that one trd75 reload was received with no apparent damage.The reload was received fully loaded with staples and the swing tab unlocked position.The reload was tested for functionality with a test device and it fired, cut, and formed all the staples as intended.After the functional test, 1 staple was noted missing along the staple line, this staple does not create a fluid path.Batch records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.The event described could not be confirmed as the reload performed without any difficulties noted.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.It should be noted that 100% inspection is performed by means of automated vision system to insure staples presence; the swing tab is present and unlocked.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|