| Model Number |
15 |
| Medical Device Problem Codes |
Failure to Charge (1085); Failure to Deliver Shock/Stimulation (1133); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/08/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's therapy pcb assembly.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report a non-critical issue with their device.There was no report of patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that the device would not detect when electrodes were connected and would not charge or shock defibrillation energy.As a result, defibrillation therapy would be unavailable, if needed.
|
| |
|
Search Alerts/Recalls
|