| Model Number |
DB-1140-S |
| Medical Device Problem Codes |
Unexpected Therapeutic Results (1631); Low impedance (2285); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
07/07/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a revision procedure, the ipg stopped functioning and had to be replaced.Once the ipg was replaced, the patient was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d10.It was reported that during a revision procedure, the ipg stopped functioning.The complaint was confirmed.Device communication could not be established due to asic u1 and battery damage.Exposing the ipg electronics to high voltage transients could cause the electrical short within the asic u1.Representative verified that patients ipg was turned off before surgery and that grounding pad was away from ipg.However, based on the signature of the damage, ipg was most likely exposed to electrocautery during the operation.The probable cause selected is unintended use error caused or contributed to event.
|
| |
|
Event or Problem Description
|
|
It was reported that during a revision procedure, the ipg stopped functioning and had to be replaced.Once the ipg was replaced, the patient was fine.
|
| |
|