| Model Number |
MMT-1715KL |
| Medical Device Problem Codes |
Visual Prompts will not Clear (2281); Moisture or Humidity Problem (2986); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had critical pump error alarm.The customer stated they were swimming at the time of event.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was received with a critical pump error (open book image) due to moisture damage on the electronic assembly.Unable to perform the functional tests, including the self test, sleep current measurement, active current measurement, rewind, prime/seating test, basic occlusion, occlusion, force sensor and displacement test due to the critical pump error.Unable to download due to download anomaly.
|
| |
|