MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON SOLERA 4.75; SCREWDRIVER Back to Search Results
Model Number 5484147
Medical Device Problem Codes Material Fragmentation (1261); Material Twisted/Bent (2981)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a consumer via field representative regarding a patient who undergone spinal therapy with an indication of eos.It was reported that during intra-op, the tip of the driver was twisted broken and remained in the patient's body.It occurred during final tightening and the fragment was difficult to retrieve because it was stuck in the set screw.Additional information received that the reported instrument was used in operation, but it was never put in the patient body.There was no patient complication as a result of this event.The broken part will be removed in 5 to 10 years, and the time is unknown at this time.No further complications were reported.
 
Additional Manufacturer Narrative
H3: analysis summary : visual inspection of part # 5484147 and lot # gb18l003 confirmed approximately 2mm of the instrument tip has been broken off, consistent with interface during usage.The remaining torx tip has been twisted.Fracture surface analysis revealed a fairly flat fracture surface and circular material flow, consistent with torsional overload.This type of damage is consistent with torsional overload.H6 : codes updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SOLERA 4.75
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key10348688
Report Number1030489-2020-01008
Device Sequence Number11362372
Product Code HXX
UDI-Device Identifier00613994591937
UDI-Public00613994591937
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5484147
Device Catalogue Number5484147
Device Lot NumberGB18L003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Initial Date Received by Manufacturer 07/14/2020
Supplement Date Received by Manufacturer12/08/2020
Initial Report FDA Received Date07/31/2020
Supplement Report FDA Received Date12/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age6 YR
Patient Weight18
-
-