| Catalog Number |
044021301 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was not provided.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is obtained, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova usa inc has been informed that, in two (2) cases, the venous line disconnected from the bcare cuvette while on bypass.Customer reattached the connection and everything worked fine the rest of the case.There is no report of any patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The complained circuit was returned to livanova.A review of the dhr did not identify any deviations, non-conformities, or material scrap/requests relevant to the reported issue.Based on livanova investigation, a potential root cause of the tubing disconnection could be an alteration of the chemical/mechanical characteristics of the tubings.Additionally, the missing tie wrap, not used by the customer, contributed to the reported issue.The risk associated to the event is acceptable.No specific action is deemed necessary livanova will keep monitoring the market.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|