Model Number 97715
Device Problems High impedance (1291); Energy Output Problem (1431); Overheating of Device (1437)
Patient Problems Pain (1994); Burning Sensation (2146)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient and manufacturing representative (rep) regarding the patient's implantable neurostimulator (ins).Information was reported that the patient reported pain and burning in the pocket.The patient stated that it sometimes gets worse when their stimulation is on, but noted that they are getting effective stimulation.There are no known factors related to this.The patient will be seen next week regarding these issues.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Additional information was received from the rep and it was reported that they met with the patient on (b)(6) 2020 to read her ins.The ins was acting properly and the leads all showed good impedance measurement except for electrode 6 (rougly 27,000 ohms) and electrode 7 (over 10,000 ohms).However, no program was using those electrodes.The patient will have an appointment with the healthcare provider (hcp) to see how to move forward.She is getting effective stimulation, just pain in the pocket site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key10349634
MDR Text Key201169470
Report Number3004209178-2020-13203
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2019
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Date Manufacturer Received08/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-