• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 SCREW 6.5MM X 20MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 SCREW 6.5MM X 20MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during incoming inspection team member found open sterile package.The sterile package was never sealed.There was no patient involvement.No additional information is available.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
Cmp-(b)(4).Reported event was confirmed by visual examination of device.Device history record (dhr) was reviewed and no discrepancies were found.The root cause of the reported event can be attributed to the operator not following instructions during manufacturing.Visual evaluation of the returned product identified that the sterile packaging pouch has not been sealed on one side.The product was likely non-conforming when it left zimmer biomet control.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 SCREW 6.5MM X 20MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10349794
Report Number0001825034-2020-03002
Device Sequence Number2655675
Product Code PBI
UDI-Device Identifier00880304527454
UDI-Public00880304527454
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number010000997
Device Lot Number6792197
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/21/2020
Supplement Date Received by Manufacturer11/03/2020
Initial Report FDA Received Date07/31/2020
Supplement Report FDA Received Date11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-