Model Number 3660 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 06/26/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.Date of event is estimated.The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
|
|
Event Description
|
It was reported the patient cannot connect to the ipg with external devices following an unrelated surgery.Surgical intervention may be pending to address the issue.
|
|
Event Description
|
Additional information indicates the ipg was explanted and replaced to address the issue.
|
|
Manufacturer Narrative
|
The reported event for no ipg communication was confirmed.It was determined the device was running the service application software.This condition is consistent with electrocautery use during surgery.Per event details, the ipg no longer communicated following an unrelated surgery.There is guidance in the clinicians manual regarding proper handling of the device when using electrocautery.As a result of this finding, actions have been taken to prevent reoccurrence.
|
|