Model Number V60 |
Device Problems
Defective Alarm (1014); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 31jul2020.
|
|
Event Description
|
The customer reported a ventilator with a primary alarm failure.The manufacturer's field service engineer confirmed the reported problem, and further confirmed this issue by observation of a speaker failure alarm during a power-on self-test in the event's history log.It was unknown if this event occurred in patient use.There was no allegation of harm associated with this event.The manufacturer's field service engineer replaced speaker assembly (#2) to resolve this reported event.
|
|
Manufacturer Narrative
|
The reported issue was confirmed an electrical open circuit in the speaker generated the reported failure of the primary alarm.
|
|