MENTOR TEXAS MENTOR SMOOTH ROUND SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Model Number |
3501490 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
06/12/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old female underwent primary breast augmentation with 575cc saline mentor smooth round spectrum breast implants and experienced deflation on the right side postoperatively.The deflation was confirmed with a physical examination.As a result, the patient underwent device removal and replacement with a 575cc saline mentor smooth round spectrum breast implant, catalog number 3501490 on (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
On september 21, 2020, the product investigation was completed.Device investigation summary: during the visual evaluation, on the posterior view of the device, some anomalies were observed consistent with crease/fold.Leak testing was performed, in accordance with mentor procedures, and no leak sites were detected in either device.As part of our quality process, the manufacturing records of this lot-serial number were reviewed and the manufacturing standards were met prior to the release of this lot.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: on (b)(6) 2020, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer's reference number: (b)(4).
|
| |
|