• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0101
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/10/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers: 3005168196-2020-01083, 3005168196-2020-01084, 3005168196-2020-01085, 3005168196-2020-01086, 3005168196-2020-01087, 3005168196-2020-01088.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the middle cerebral artery.During the procedure, the patient¿s vessel was perforated using a non-penumbra guidewire.The physician attempted to use penumbra smart coils (smart coils) to stop the bleeding; however, six smart coils would not advance past their initial positions within the introducer sheaths.Therefore, the smart coils were not used in the procedure.The next three smart coils were implanted in the target location using the penumbra smart coil detachment handle (handle).While attempting to implant the next smart coil using the handle, it would not detach.The physician attempted to manually detach the smart coil without success; therefore, this smart coil was removed from the patient.The procedure was completed using five smart coils and the same handle.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was separated approximately 0.3 cm on the proximal end of the pusher assembly.Kinks and bends were present along the length of the pusher assembly.The pull wire was within the pusher assembly distal detachment tip (ddt).The embolization coil was intact with the pusher assembly and had offset coil winds.The introducer sheath was not returned for evaluation.Conclusions: evaluation of the first smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If this occurs, resistance will likely be experienced while advancing the device.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.Then the device was able to advance through its introducer sheath and a demonstration microcatheter without an issue.Evaluation of the second smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If this occurs, resistance will likely be experienced while advancing the device.Forcefully advancing the device against this resistance may have contributed to the kinks which were observed on the pusher assembly.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.Then the device was able to advance through its introducer sheath and a demonstration microcatheter with resistance due to the kinks on the pusher assembly.Evaluation of the third smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If this occurs, resistance will likely be experienced while advancing the device.Forcefully advancing the device against this resistance may have contributed to the kinks which were observed on the pusher assembly.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.Then the device was able to advance through its introducer sheath and a demonstration microcatheter with resistance due to the kinks on the pusher assembly.Evaluation of the seventh smart coil confirmed that the embolization coil was unable to detached from its pusher assembly.Further evaluation of the device revealed that the pet lock was separated; however, the pull wire was not retracted out of the ddt.If pull wire is not retracted from the ddt, the embolization coil will not detach from the pusher assembly.During functional testing, the embolization coil could not be detached with a demonstration handle.Therefore, the smart coil was attempted to detach manually.While retracting the pull wire from the proximal end of the pusher assembly, resistance was encountered, and the pull wire was unable to retract from the ddt.Therefore, the smart coil could not be manually detached.Further evaluation of the device also revealed kinks and bends on the pusher assembly and offset coil winds on the embolization coil.The kinks and bends may have contributed to the resistance during the functional test.The offset coil winds were likely incidental to the complaint.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2020-01083; 3005168196-2020-01084; 3005168196-2020-01085; 3005168196-2020-01086; 3005168196-2020-01087; 3005168196-2020-01088.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10354070
Report Number3005168196-2020-01089
Device Sequence Number3064957
Product Code HCG
UDI-Device Identifier00814548015590
UDI-Public00814548015590
Combination Product (Y/N)Y
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/10/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/03/2025
Device Model Number400SMTHXSFT0101
Device Catalogue Number400SMTHXSFT0101
Device Lot NumberF97660
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/10/2020
Supplement Date Received by Manufacturer09/22/2020
Initial Report FDA Received Date08/01/2020
Supplement Report FDA Received Date10/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-