• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0101
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/10/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2020-01083, 3005168196-2020-01084, 3005168196-2020-01085, 3005168196-2020-01086, 3005168196-2020-01088, 3005168196-2020-01089.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the middle cerebral artery.During the procedure, the patient¿s vessel was perforated using a non-penumbra guidewire.The physician attempted to use penumbra smart coils (smart coils) to stop the bleeding; however, six smart coils would not advance past their initial positions within the introducer sheaths.Therefore, the smart coils were not used in the procedure.The next three smart coils were implanted in the target location using the penumbra smart coil detachment handle (handle).While attempting to implant the next smart coil using the handle, it would not detach.The physician attempted to manually detach the smart coil without success; therefore, this smart coil was removed from the patient.The procedure was completed using five smart coils and the same handle.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10354072
Report Number3005168196-2020-01087
Device Sequence Number19599257
Product Code HCG
UDI-Device Identifier00814548015590
UDI-Public00814548015590
Combination Product (Y/N)Y
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/10/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/03/2025
Device Model Number400SMTHXSFT0101
Device Catalogue Number400SMTHXSFT0101
Device Lot NumberF97660
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2020
Initial Report FDA Received Date08/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-