| Model Number |
MMT-751LNAS |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a compromised force sensor system alarm.The customer stated the insulin pump was resetting so they were unable to use the insulin pump.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test and excessive no delivery test or verify a33 alarm due to motor error alarms.Device received with cracked belt clip slot and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.(b)(4).
|
| |
|