• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAP
Medical Device Problem Code Failure to Cycle (1142)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump received compromised forced sensor alarm during fill tuning.Drive support cap was normal.Customer was able to rewind the insulin pump.No harm requiring medical intervention was noted.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the displacement test but a33 alarm during the basic occlusion test due to protruded drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to a33 alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10355661
Report Number2032227-2020-156695
Device Sequence Number2841578
Product Code OYC
UDI-Device Identifier00643169513853
UDI-Public(01)00643169513853
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAP
Device Catalogue NumberMMT-523LNAP
Device Lot NumberA4523LNAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2020
Initial Date Received by Manufacturer 07/28/2020
Supplement Date Received by Manufacturer09/14/2020
Initial Report FDA Received Date08/03/2020
Supplement Report FDA Received Date09/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age21 YR
Patient Weight64
-
-