| Model Number |
MMT-751NAP |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the retainer ring had crack.Customer reported that there was physical damage to the insulin pump's retainer ring and the reservoir was not able to lock in place when inserted into insulin pump.No harm requiring medical intervention was reported.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement.Device received with complete broken reservoir tube lip, cracked belt clip slot and missing reservoir tube o-ring.The p-cap not locks into the reservoir compartment properly due to completely broken reservoir tube lip noted.
|
| |
|