| Model Number |
26665 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was received with the stent in the correct place on the delivery system.The stent could not be fully deployed due to a severe kink and complete break in the outer shaft of the device.A visual and tactile inspection identified a complete break of the outer shaft located approximately at the guidewire port.A severe kink on the inner shaft was also identified at the same location.This type of damage is consistent with excessive force being applied to the device.A visual and microscopic examination identified no issues with the tip of the device.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 23jul2020.It was reported that stent deployment failure occurred.A 10x31,5.9f,135cm carotid wallstent was advanced for treatment but failed to deploy.The device was removed and the procedure was completed with another of the same device.No patient complications were reported.However, returned device analysis revealed shaft detached/separated.
|
| |
|