| Model Number |
PROPAQ MD |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device failed self-test for defib function.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's report was observed upon power up; however, after disassembling the device to determine root cause, the report was no longer seen.The device was put through extensive testing including bench handling, power cycling, and defib functional stress testing without duplicating the report.The processor/bridge/pace board was replaced and scrapped as a precaution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer was contacted for return of the suspect product.The product has not been returned to zoll for evaluation.
|
| |
|