| Model Number |
SPMB8 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
06/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/ non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.Patient identifier unknown / not provided.Patient sex unknown / not provided.Weight unknown / not provided.Pma/510(k) number: k011245 and k002188.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant did not achieve primary stability during implant placement due to bone density.
|
| |
|
Search Alerts/Recalls
|