| Model Number |
MICL13.7 |
| Medical Device Problem Codes |
Off-Label Use (1494); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon was loading a 13.7mm micl13.7 implantable collamer lens, -16.00 diopter, and noted a nick in the lens.There was no patient contact and the backup lens was implanted.The problem was resolved and the patient is doing well.The cause of the event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned in liquid in lens vial.Visual inspection found the lens haptic torn.H6 - device code: 1494 - this lens model is contraindicated for patients with age less than that of 21 years.Claim # : (b)(4).
|
| |
|