| Model Number |
1217-01-058 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Codes |
Pain (1994); No Code Available (3191)
|
| Date of Event |
07/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Clinical adverse event received for implant fracture - s-rom stem.Event is serious and is considered severe.Event is definitely related to device and is definitely not related to procedure.Date of implantation: (b)(6) 2008.Date of event (onset): (b)(6) 2020 (right hip).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Examination of provided x-ray images confirms the reported allegation.The position of the acetabular cup is more vertical than would be recommended by depuy.Cup placement is determined by the implanting surgeon and is not considered a product problem.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Corrected: h6 code 3191 (patient) from absence of treatment to device revision or replacement.
|
| |
|