• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX ADVANCE BR W/OC&ND; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX ADVANCE BR W/OC&ND; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 222299
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event or Problem Description
It was reported by the sales representative via phone that two 5.5 healix advance br w/oc&nd anchors broke when inserted.It was reported the third device worked perfectly.No harm to patient.No significant delay in surgery time.Additional information provided by the sales representative reported only 1 device will be returned for evaluation because the other device was discarded.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported that during the surgery of rotator cuff suture, the anchor was broken off.The complaint device was received and evaluated.Visual observation confirms the distal tip of the anchor was broken and was found detached form the shaft, but is held in place by suture.In addition, the device presented evidence of biological residues in the anchor, handle and in the sutures.When observed under magnification, no structural anomalies were observed.The complaint can be confirmed.The possible root cause for the reported failure can be associated with off axis insertion, levering during insertion, hard bone quality or using incorrect instrumentation for preparing bone hole.However, this cannot be conclusive determined.A manufacturing record evaluation was performed for the finished device [6l77852] number, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX ADVANCE BR W/OC&ND
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10359118
Report Number1221934-2020-01988
Device Sequence Number3462611
Product Code MAI
UDI-Device Identifier10886705021352
UDI-Public10886705021352
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K120078
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number222299
Device Catalogue Number222299
Device Lot Number6L77852
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2020
Initial Date Received by Manufacturer 07/17/2020
Supplement Date Received by Manufacturer09/07/2020
Initial Report FDA Received Date08/03/2020
Supplement Report FDA Received Date09/08/2020
Patient Sequence Number1
-
-