DEPUY ORTHOPAEDICS INC US 28MM PINN ESC CONS RING INSRTR; HIP INSTRUMENTS : INSERTION DEVICES Back to Search Results
Model Number 2218-00-028
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Deformation (2976); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the ring inserter no longer fits symmetrically around ring to impact the ring.It looks bent.It has been requested to be replaced per the surgeon.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
28MM PINN ESC CONS RING INSRTR
Common Device Name
HIP INSTRUMENTS : INSERTION DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10359173
Report Number1818910-2020-17404
Device Sequence Number11426786
Product Code HWR
UDI-Device Identifier10603295099529
UDI-Public10603295099529
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2218-00-028
Device Catalogue Number221800028
Device Lot NumberSO2001520
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2020
Initial Date Received by Manufacturer 07/26/2020
Supplement Date Received by Manufacturer08/01/2020
08/07/2020
Initial Report FDA Received Date08/03/2020
Supplement Report FDA Received Date08/06/2020
08/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-