|
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, weight, race, ethnicity, and medical history were not provided.Procode is krd/hcg.[conclusion]: the healthcare professional reported that during the embolization targeting a 12mm x 8mm middle cerebral artery (mca) aneurysm, while deploying the 7mm x 13.9cm micrusframe 10 coil (mfr100713 / l15689), the last loop of the coil was pushing on the concomitant sl-10® microcatheter (stryker) causing the microcatheter to retreat.There was no difficulty with positioning the coil in the target lesion, there was no issue with the coil conforming to the aneurysm wall.It was reported that the coil was withdrawn and repositioning two times.The coil was removed still attached to the delivery system.No damage was noted on the coil when it was removed.The embolic coil and its respective device positioning unit (dpu) did not protrude from the coil introducer.The physician had the same experience with three other coils: another 7mm x 13.9cm micrusframe 10 coil from the same lot (l15689), a 4mm x 10cm galaxy g3 xsft (glx120410 / l16657), and a 6mm x 11.9cm micrusframe 10 coil (mfr100611 / l16500).The physician switched to axium coils (medtronic) and was able to deliver the axium coils using the same sl-10® microcatheter to complete the procedure.There was no report of any significant procedural delay; there was no report of any patient adverse event or complication.The complaint coil was returned for evaluation and analysis.The investigational finding is documented below.Investigation summary: the non-sterile 7mm x 13.9cm micrusframe 10 coil was received contained in a pouch.Visual inspection was performed.There was no appearance of damage observed.The returned device underwent a microscopic inspection.The marker band was found at 49cm from the distal end; this is within specification.The embolic coil was observed in slightly kinked condition in two areas.No other damage was observed under the magnification of the microscopic inspection.A review of manufacturing documentation associated with this lot (l15689) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.The complaint documented that the last loop of the coil was pushing on the concomitant microcatheter causing it to retreat.The embolic coil of the returned device was observed in kinked condition.The slight kinked areas on embolic coil are the likely contributing factors to the reported issue that the coil was ¿pushing¿ on the concomitant microcatheter causing the microcatheter to retreat during the procedure.The issue was confirmed based on the observed kinked areas on the embolic coil of the returned device.Coil kinking is a known potential issue associated with the use of the device.The instruction for use provides proper handling instructions for the device to prevent issues such as kink from occurring.The kinked condition observed on the embolic coil of the returned device was not originally reported with the complaint.The exact cause of the observed kinked condition could not be conclusively determined; however, it may have been the result of inadvertent force that may have been applied during the procedural device handling.Based on the review of the manufacturing documentation, there is no indication that the event is related to the device manufacturing process.In addition, devices undergo 100% inspection at final assembly for the condition of the embolic coil.Thus, it is not likely that the 7mm x 13.9cm micrusframe 10 coil left the manufacturing facility with the embolic coil in the observed kinked condition.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This is one of four (4) product events involved with the literature publication.The associated manufacturer report numbers are: 3008114965-2020-00321, 3008114965-2020-00322, and 3008114965-2020-00323.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
The healthcare professional reported that during the embolization targeting a 12mm x 8mm middle cerebral artery (mca) aneurysm, while deploying the 7mm x 13.9cm micrusframe 10 coil (mfr100713 / l15689), the last loop of the coil was pushing on the concomitant sl-10® microcatheter (stryker) causing the microcatheter to retreat.There was no difficulty with positioning the coil in the target lesion, there was no issue with the coil conforming to the aneurysm wall.It was reported that the coil was withdrawn and repositioning two times.The coil was removed still attached to the delivery system.No damage was noted on the coil when it was removed.The embolic coil and its respective device positioning unit (dpu) did not protrude from the coil introducer.The physician had the same experience with three other coils: another 7mm x 13.9cm micrusframe 10 coil from the same lot (l15689), a 4mm x 10cm galaxy g3 xsft (glx120410 / l16657), and a 6mm x 11.9cm micrusframe 10 coil (mfr100611 / l16500).The physician switched to axium coils (medtronic) and was able to deliver the axium coils using the same sl-10® microcatheter to complete the procedure.There was no report of any significant procedural delay; there was no report of any patient adverse event or complication.The product was returned for evaluation; during the microscopic inspection of the device, the embolic coil was observed in kinked condition.Based on the product analysis completed on (b)(6) 2020, this event has been deemed mdr reportable as a ¿malfunction.¿.
|