| Model Number |
9-ITV09F45/80 |
| Medical Device Problem Code |
Biocompatibility (2886)
|
| Health Effect - Clinical Code |
Thrombus (2101)
|
| Date of Event |
07/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a 18mm amplatzer septal occluder(aso) was selected for implant in the patient after measuring the size with a sizing balloon.During the procedure the aso was attempted to be implanted, but the physician decided that the aso was under-sized.Thrombus had formed in the patient and adhered to the delivery cable during the procedure.The physician was concerned about the thrombus adhering on the delivery cable and decided to remove the 18mm aso and the 9f 45/80 amplatzer torqvue delivery system from the patient.The devices were replaced with a 20mm amplatzer septal occluder(aso) and a 10f 45/80 amplatzer torqvue delivery system to continue the procedure.The 20mm aso was successfully implanted with no adverse consequences to the patient.It took time in the attempt to deploy the first 18mm aso in the patient's defect, so the delivery cable and the sheath were in the patient longer than usual procedures.However, the delay was not significant and the patient remained hemodynamically stable during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of thrombus in the delivery cable was reported.The investigation found that residue consistent with blood was present on the delivery cable when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
|
| |
|
Search Alerts/Recalls
|