Catalog Number 71692802
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/03/2020
Event Type  Injury  
Event Description
It was reported that during set up or inspection monitor does not work properly.Procedure was completed with a different technique.No patient harm or delay was reported.
 
Manufacturer Narrative
Results of investigation: it was reported that during set up or inspection monitor does not work properly.Procedure was completed with a different technique.No delay reported.The affected complaint device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of manufacturing records did not reveal any deviation from the standard manufacturing processes.A review of the complaint history on the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship, if any, between the device and the reported incident could not be corroborated.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.Some potential causes could include but are not limited to hardware failure or electrical interference.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMA INTERFACE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10359984
MDR Text Key201494977
Report Number1020279-2020-03667
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K170280
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number71692802
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/10/2020
Initial Date FDA Received08/03/2020
Supplement Dates Manufacturer Received03/17/2021
Supplement Dates FDA Received04/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-