• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES Back to Search Results
Model Number MAJ-1516
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/10/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device has not returned to olympus medical systems corp.(omsc) for evaluation.There were no further details provided.If significant additional information is received, this report will be supplemented.
 
Event or Problem Description
During inspection prior to use, the device's lock lever didn't fit in with automated endoscope reprocessor.The customer used another connecting tube to complete the reprocessing.There was no report of patient injury associated with the event.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc could not review the manufacturing history (dhr) of the subject device because the lot number of the device is unknown.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the investigation result by omsc there was the possibility that the reported phenomenon was attributed to the accidental external force applied to the lock lever such as during transportation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING TUBE
Common Device Name
ENDOSCOPE REPROCESSOR ACCESSORIES
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10360192
Report Number8010047-2020-05215
Device Sequence Number19555204
Product Code FEB
UDI-Device Identifier04953170234057
UDI-Public04953170234057
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K120357
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberMAJ-1516
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/13/2020
Supplement Date Received by Manufacturer12/10/2020
Initial Report FDA Received Date08/03/2020
Supplement Report FDA Received Date12/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-