| Model Number |
MMT-751NAP |
| Medical Device Problem Codes |
Mechanical Problem (1384); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump had motor error alarm and no delivery alarm.The customer stated that the drive support cap was normal.Customer reported that they were able to clear and rewind the insulin pump.Motor error alarm was again happened after troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump will be returned for nalysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery test.No excessive no delivery or occlusion alarm noted.No motor error alarm noted.Motor passed motor test.(b)(4).
|
| |
|
Search Alerts/Recalls
|