Catalog Number 03666794001 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reporter's product was requested for investigation and replacement product was sent to the reporter.Occupation: the occupation is lay user/patient.
|
|
Event Description
|
The initial reporter stated that the display of his coaguchek xs meter serial number (b)(4) is faded and he can not read it.The reporter performed a display check and some of the segments were missing from the result field of the display.There may be a possibility of a result misinterpretation.The reporter stated he could not see anything on the display when looking at it from the top.He stated that he needs to turn it to the side in order to see anything.The battery compartment of the meter was clean and dry.There was no evidence of burning, melting, or smoking.The buttons of the meter were not stuck or damaged.
|
|
Manufacturer Narrative
|
The customer's meter was received for investigation.The meter did not turn on during investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The observed contamination can lead to the reported display issue.The root cause was determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d10 and h3 were updated.
|
|