Catalog Number 03666794001
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/13/2020
Event Type  malfunction  
Manufacturer Narrative
The reporter's product was requested for investigation and replacement product was sent to the reporter.Occupation: the occupation is lay user/patient.
 
Event Description
The initial reporter stated that the display of his coaguchek xs meter serial number (b)(4) is faded and he can not read it.The reporter performed a display check and some of the segments were missing from the result field of the display.There may be a possibility of a result misinterpretation.The reporter stated he could not see anything on the display when looking at it from the top.He stated that he needs to turn it to the side in order to see anything.The battery compartment of the meter was clean and dry.There was no evidence of burning, melting, or smoking.The buttons of the meter were not stuck or damaged.
 
Manufacturer Narrative
The customer's meter was received for investigation.The meter did not turn on during investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The observed contamination can lead to the reported display issue.The root cause was determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d10 and h3 were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10361389
MDR Text Key201573002
Report Number1823260-2020-01883
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2020
Date Manufacturer Received07/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-