| Model Number |
MMT-751NAB |
| Medical Device Problem Codes |
No Display/Image (1183); Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the bolus wizard did not give enough insulin even when there was insulin in the reservoir.The insulin pump was not giving correct amount of insulin.They also reported display issue.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display and constant tone due to moisture damage on the electronics assembly.Unable to perform all functional testing including the operating currents, a21 error test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test or verify ender delivery anomaly due to blank display.Device received with moisture damage motor, vibrator and battery tube assembly.
|
| |
|