| Model Number |
MMT-723RNAP |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump received reset alarm and unexpected power loss alarm.Customer was able to clear the alarm.Customer was able to rewind the insulin pump.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test and a21 alarm test.No unexpected a21 alarm anomalies noted.No unexpected reset alarm noted.(b)(4).
|
| |
|