| Model Number |
3L92509 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Pain (1994); Osteolysis (2377); No Code Available (3191)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient had hip surgery many years ago in different hospital.Hybride construct , later revised tot cementless construct( cemented cup replaced) earlier this year patient came back with pain in hip and x-ray showed osteolysis proximaly around hip prothesis.Patient cam to the hospital with pain.No delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.
|
| |
|
Search Alerts/Recalls
|