Model Number 3L92509
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377); No Code Available (3191)
Date of Event 01/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient had hip surgery many years ago in different hospital.Hybride construct , later revised tot cementless construct( cemented cup replaced) earlier this year patient came back with pain in hip and x-ray showed osteolysis proximaly around hip prothesis.Patient cam to the hospital with pain.No delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 STD SIZE 9
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key10362321
Report Number1818910-2020-17454
Device Sequence Number951175
Product Code KWA
UDI-Device Identifier10603295168751
UDI-Public10603295168751
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K042992
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3L92509
Device Catalogue Number3L92509
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/22/2020
Supplement Date Received by Manufacturer10/13/2020
Initial Report FDA Received Date08/04/2020
Supplement Report FDA Received Date10/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP ACETABULAR LINERS; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP ACETABULAR LINERS; UNKNOWN HIP FEMORAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
-
-