| Model Number |
9438-05 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Per the tandem user guide: "the transmitter battery will last at least 90 days.Once you see the low transmitter battery alert, replace the transmitter as soon as possible.No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that intermittent loss of connection issues occurred between the transmitter and the pump.Reportedly, the transmitter had been in use longer than 90 days; customer support instructed customer to change transmitter.No additional patient information was available.
|
| |
|
Search Alerts/Recalls
|