| Model Number |
MMT-751LNAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Battery Problem (2885); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had motor error, had rejected new batteries and the battery lasted only for a day since a month or two.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed all functional testing including the displacement, operating current test, a21 error test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test.No motor error alarm, no failed battery test alert or unexpected battery power loss anomaly during test noted.Motor passed motor test.(b)(4).
|
| |
|