• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3231-40Q
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
Following the advisory for premature battery depletion with implantable cardioverter defibrillator, although there was no eri alert or allegation of premature battery depletion, the device was explanted prophylactically.
 
Additional Manufacturer Narrative
Interrogation of the device revealed the device was at eri when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Based on this information, there was no evidence of premature depletion.The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on 11 october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10365388
Report Number2017865-2020-10511
Device Sequence Number16895244
Product Code NIK
UDI-Device Identifier05414734504393
UDI-Public05414734504393
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2012
Device Model NumberCD3231-40Q
Device Catalogue NumberCD3231-40Q
Device Lot Number3253380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/16/2020
Supplement Date Received by Manufacturer12/18/2020
Initial Report FDA Received Date08/04/2020
Supplement Report FDA Received Date12/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-